MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
23
23
Note: When the battery is discharged, the status indicator will display
“
X
”
. Even if the new
battery is replaced, the status indicator still displays
“
X
”
. Please contact qualified
service personnel or your local supplier.
Note: Self test is not able to determine if the battery currently inserted in the AED are
within their use by date. User must remember to check the use by date on the
battery and standby life on the battery regularly.
Note: During the self test, the temperature status is displayed when the ambient
temperature is out of range.
Note: The AED that is activated the record function will do the self test for record function.
Note: AED will always do the self test for the pads connection. If the pads are
disconnected to the AED, the pads status indicator is displayed.
Note: The AED that is activated the pads quality function will do the self test for use by
date of the pads, whether used, dryness and damage.
Note: The AED that is activated the CPR Feedback function will do the self test for CPR
Feedback function.
Battery Insertion Self test
When the battery is replaced, the AED au
tomatically starts the battery insertion self test. After battery insertion self test is
completed,
the AED sounds voice prompt “Unit ok”, the status indicator displays “O” and
power of the AED is automatically turned off. If the battery insertion self test is not
completed successfully, the AED sounds voice prompt
“
Unit fail
”
and the status indicator
displays
“
X
”
. If the AED does not function properly, do not use the AED. Instead, contact
qualified service personnel or your local supplier.
You can also skip the battery insertion self test, try following procedure.
Open the cover of the AED, and pushing the power button.
Note: Self test is not able to determine if the battery currently inserted in the AED are
within their use by date. User must remember to check the use by date on the
battery and standby life on the battery regularly.
Note: The AED that is activated the pads quality function will do the self test for use by
date of the pads, whether used, dryness and damage. User must remember to
check the pads condition regularly.
Note: The internal battery is used to perform self test and real time clock and it could
operate for at least 10 years. However, please note that this period could vary
slightly depending on storage condition and use of environment. If the operating
time has passed, the battery status disappears and
“
X
”
appears on the status
indicator. In this case, please contact qualified service personnel or your local
supplier.