![Mediana A16M-DF-0E Скачать руководство пользователя страница 18](http://html1.mh-extra.com/html/mediana/a16m-df-0e/a16m-df-0e_operators-manual_1760607018.webp)
MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
14
14
Soft Carry Case
The AED should be protected by the soft carry case during the move.
The soft carry case has been designed to allow the AED not to move in the soft carry
case by using the AED own handle. The user can check the status indicator of AED
without having to open the carry case. The paper with contact information of the nearest
emergency medical services can be inserted to the clear cover. The soft carry case has
the pocket on the rear side of the carry case for the manual and spare pads, spare
battery and the CPR Feedback module (option).
SD card
The SD card is inserted into the SD card port on the AED
’
s bottom panel as described
below. The SD card is used to record the history of the AED performance and to update
the AED firmware. The recorded history in the SD card can be checked by the HeartOn
AED Event Review Software. If you want to use the SD card to use the HeartOn AED
Event Review Software or to update the AED firmware, please contact qualified service
personnel or your local supplier.
1. When the AED is turned on, turn off the AED by pushing the power button.
2. Remove the battery from the bottom of the AED.
3. Check the proper orientation of the SD card and insert the SD card into the SD card
port.
4. Reseat the battery in the bottom of the AED.
5. If necessary to update the AED through SD card, turn on the AED by pushing the
power button.
6. After complete the update, automatically turn off itself. Close the cover again.
Note: Only micro SD cards supplied by Mediana should be used.
Event Data
The event data are stored in the SD card. The event data can be read by HeartOn AED
Event Review Software.
Note: When the AED does not have an SD card in it or your SD card is unreadable,
corrupted, damaged or has some error, the event data are stored in the internal
memory as the LED indicator and buzzer sound will go on.
In case the SD card has a problem, buzzer will sound 4 times after the CPR Action
Icon LED lights up. {Repeat twice}
AED records up to 4 times (1 time: from power ON to power OFF) of operation data in
the flash memory. From the fifth operation data will be stored after the oldest data is
deleted.
The maximum data that can be stored at one time is ECG (6488 sec), HR (37 times of
shock advised, 74 times of no shock advised), Trend (37 times of shock advised, 74
times of no shock advised), and Temperature (36 times). If any of the above four data is
full during a progress once, the operation data is no longer stored.