Step 3
Semi-Auto
“Press the flashing button
now.
”
“Shock delivered
.
”
or
“Press the flashing button now
.
”
“
Shock button not pressed.
”
When the ECG rhythm is changed to a non-shockable rhythm before the shock button
is pressed or when the pads are disconnected, the following voice prompt will be
emitted:
“Shock cancelled
.
”
Note: If the shock button is not pressed for 10 seconds after the voice prompt
“
Press the
flashing button now.
”
, the same instruction will be emitted one more time. If the
shock button is not pressed for more than 20 seconds after the initial instruction,
the AED performs the CPR process after the voice prompt
“
Shock cancelled
”
is
emitted.
Note: If the shock cancel (when the pad connector is disconnected, the patient ECG,
impedance, or mode is changed) has been occurred 3 times, the AED is
performing the CPR process after the voice prompt
“
Shock cancelled
”
is emitted.
Fully-Auto
“Shock
will be delivered.
”
“
Do not touch the patient.
”
“
Three, Two, One.
”
“
Shock delivered.
”
or
“
Shock cancelled.
”
Figure 14. Action Icon
–
Step 3
(Left: Chest type
∙
A16-DS, A16-DF (
ⓒ,ⓔ
), Right: Icon type
∙
A16-GS, A16-GF(
ⓒ,ⓔ
)