MEDIANA
HEART
ON
AED A16
Automated External Defibrillator
2
2
Proper placement of the pads is critical. Strict observance of pad
positioning instructions, as indicated on the labeling and in training, is
essential. Care must be taken to ensure pads are adhered to the patients'
skin properly. Air pockets between the adhesive pad and skin must be
eliminated. Failure in pad adhesion may hinder effectiveness of therapy
or cause excessive skin burns to the patient if a shock is applied.
Reddening of the skin may appear after use, this is normal.
The battery of AED is not rechargeable. Do not try to recharge, open,
crush, or burn the battery, or it may explode or catch fire.
Do not allow the pads to contact other electrodes or metal parts that are
in contact with the patient. Such contact can cause patient skin burns
during defibrillation and may divert defibrillating current away from the
heart.
Pay attention to possibility of contact with conduction part of electrode,
lead line, cable connector, other patient installation part for patient
safety.
Do not use this AED near or within puddles of water.
Do not reuse electrodes to many patients.
Use the AED or accessories only as described in this manual. Improper
use of the AED can cause death or injury.
Do not use or place the AED in service if the status indicator of AED
displays
“
X
”
. Contact your local distributor or Mediana service team.
Keep batteries dry and away from any heat sources (including direct
sunlight). If you see any damage or leakage, do not allow the liquid to
come in contact with your skin or eyes. If contact has been made, wash
the affected area with plenty of water and seek medical advice
immediately.
The AED contains an automatic disarm of the stored energy. If the
operator has not delivered the energy to a patient, an internal timer will
disarm the stored energy. This stored electrical energy can potentially
cause death or injury if discharged improperly. Follow all instructions in
this manual.
Check the status Indicator periodically. If the status Indicator displays
“X”
, a problem has been detected. Manage it for no problem in case of
emergency.
As with any medical equipment, carefully route patient cabling to reduce
the possibility of patient entanglement or strangulation.