Automated External Defibrillator
46
46
EMI (Electromagnetic Interference)
Keep patients under close surveillance during delivering a shock. It is
possible, although unlikely, that radiated electromagnetic signals from
sources external to the patient and the AED can cause inaccurate
measurement readings. Do not rely entirely on the AED readings for
patient assessment.
It is possible that any radio frequency transmitting equipment and other
nearby sources of electrical noise may result in disruption in the AED
operation.
It is possible, although unlikely, that large equipment using a switching
relay for its power on/off may affect the AED operation. Do not operate
the AED in conjunction with electrocautery or diathermy equipment or in
such environments.
This AED has been tested and found to comply with the limits for Medical devices to the
IEC60601-1-2, and the Medical device Directive 93/42/EEC. These limits are designed
to provide reasonable protection against harmful interference in a typical medical
installation.
However, because of the proliferation of radio-frequency transmitting equipment and
other sources of electrical noise in health care environments (such as electrosurgical
equipment, cellular phones, mobile two-way radios, electrical appliances, and high-
definition television), it is possible that high levels of such interference due to close
proximity or strength of a source may affect AED operation.
The AED is designed for use in environments in which the signal can be
obscured by electromagnetic interference. During such interference,
measurements may seem inappropriate or the AED may not seem to
operate correctly.
The AED disruption may be indicated by erratic readings, cessation of operation, or
other incorrect functioning. If this occurs, survey the site to determine the source of this
disruption. Try the following actions to see if they eliminate the disruption:
Turn equipment in the vicinity off and on to isolate the offending equipment.
Reorient or relocate the interfering equipment.
Increase the separation between the interfering equipment and this equipment.
The AED generates, uses, and can radiate radio frequency energy. If the AED is not
installed and used in accordance with these instructions, the AED may cause harmful
interference with other devices in the vicinity.
If assistance is required, contact qualified service personnel or your local supplier.
Obtaining Technical Assistance
For technical information and assistance, or to order the AED service manual, call your
local supplier. The service manual provides information required by qualified service
personnel or your local supplier when servicing the AED.