Manufacturer
’
s Declaration
For best product performance and measurement accuracy, use only
accessories supplied or recommended by Mediana. Use accessories
according to the manufacturer’s directions for use and your facility’s
standards. The use of accessories, transducers, and cables other than
those specified may result in increased emission and/or decreased
immunity of the AED.
Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the AED, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this
equipment could result.
The AED is suitable for use in the specified electromagnetic environment. The
customer and/or user of the AED should assure that it is used in an electromagnetic
environment as described below;
Table 5. Electromagnetic Emissions (IEC60601-1-2)
Emission Test
Compliance
Electromagnetic Environment
RF emission
CISPR 11
Group 1
The AED must emit electromagnetic energy
in order to perform its intended function.
Nearby electronic equipment may be
affected.
RF emissions
CISPR 11
Class B
The AED is suitable for use in all
establishments.
Table 6. Electromagnetic Immunity (IEC60601-1-2)
Immunity Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic
Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floor should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30 %.
Power frequency
(50 Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
It may be necessary to position
the AED further from the
sources of power frequency
magnetic fields or to install
magnetic shielding. The power
frequency magnetic field
should be measured in the
intended installation location to
assure that it is sufficiently low.
Note: UT is the AC mains voltage prior to application of the test level.