M
EDEX
Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 9
Syringe Manufacturers & Standard Sizes
Below is a table of syringe manufacturers for which the
Medfusion/or 3500
has been designed, and the specific
syringe sizes programmed into the pump database for use with the pump.
Manufacturer
Syringe Sizes
B. Braun
Omnifix
®
5, 10, 20, and 50 ml Luer Lock
Perfusor
®
20 and 50ml Luer Lock
Becton-Dickinson Luer Lok
®
B-D 3, 5, 10, 20, 30, and 60 ml Luer Lock
B-D 1ml Luer Lock, and 1ml Tuberculin
Becton-Dickinson Glass
1, 2.5, 5 and 10 ml Luer Slip
Monoject
®
3, 6, 12, 20, 35, and 60 ml Luer Lock
1ml Slip
Terumo
®
3, 5, 10, 20, 30, and 60 ml Luer Lock
1ml Slip
AstraZeneca
®
50 ml pre-filled Diprivan
®
Luer Lock
Warning
!
Critical Volume
The critical volume (for
volume limit
) which could be infused in the event of a single
point positional failure in this infusion pump is 1/100
th
of the syringe fill volume.
At
any rate
the critical volume (for
rate
) in this infusion pump is 1/400
th
of the syringe
fill volume.
T
RADEMARK
I
NFORMATION
Below is a table of registered trademarks and the manufacturers owning them:
Trademark
Owner / Manufacturer
“Omnifix” and “Perfusor”
B.Braun Medical AG
“B-D”
Becton Dickinson and Co
“Monoject”
Sherwood Medical
“Diprivan”
AstraZeneca Pharmaceutical