Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 16
Alarm Message
Priority
Remedy
Syringe Barrel Not in Place
High
The syringe is not loaded correctly. Make sure the syringe barrel clamp is
correctly positioned and is holding the syringe in place. Proceed with the
infusion.
Syringe Flange Not in Place
High
The syringe is not loaded correctly. Make sure the syringe flange is
properly inserted into the flange clip. Proceed with the infusion.
Syringe Plunger Not in Place
High
The syringe is not loaded correctly Make sure the syringe plunger head is
properly inserted. Proceed with the infusion.
Syringe Does Not Match
Library Entry
Limit
Program the delivery for the type of syringe you are actually using. Then
proceed with infusion setup and start it.
Syringe Empty
High
The pump has halted infusion at empty. You may either:
Prepare and load another syringe, then begin another infusion.
Press the
Start
button to override the empty alarm and deliver any
remaining fluid in the syringe.
Return to the S
ELECT THE
M
ODE
screen, where no alarms sound.
Turn off the pump.
Syringe Empty
−
Manual
High
The pump has halted infusion after overriding empty alarm. You may
either:
Prepare and load another syringe, then begin another infusion.
Return to the S
ELECT THE
M
ODE
screen, where no alarms sound.
Turn off the pump.
Syringe Near Empty
Low
The pump is telling you the syringe will soon be empty.
System Advisory
−
Maintenance is
Recommended
Low
As soon as possible, remove pump from service for testing and checkout by
a trained biomedical service technician.
System Advisory
−
Set Time
and Date
Low
As soon as possible, remove pump from service for testing and checkout by
a trained biomedical service technician.
System Advisory
−
xx Data
Corrupted
Low
Infusion will continue using backup copy of
xx data
. As soon as possible,
remove pump from service for testing and checkout by a trained biomedical
service technician.
System Failure
−
aa bb cc
High
Press
Power
button to turn off pump. Immediately remove pump from
service for repair by a trained biomedical service technician.
Below Recommended Rate
Limit
Reminds user that the programmed rate is below that recommended for the
syringe size - no remedy required