Page 30
Intermittent Volume / Time
This
Intermittent Volume / Time Mode
lets you specify a delivery volume for a specific delivery time, after which the
delivery stops for a programmed interval … then the pattern recycles. In other words, the pump delivers for a set
period, then pauses a set period, then infuses for a set period, and so on.
•
The
delivery stop interval
equals “
Time Between Starts
” minus “
Delivery Time
.”
•
The pattern
repeats
at “Time Between Starts” interval.
Below are the steps for programming an Intermittent Volume / Time infusion. Select Intermittent Volume / Time
from the S
ELECT THE
M
ODE
screen. This calls the V
OLUME TO
I
NFUSE
screen.
1.
Use the
number
buttons to set the volume. Volume here is limited to the maximum fill volume for the
syringe size, and minimum volume is 1/600
th
of syringe size.
2.
When ready, press the
Enter
button to accept the Volume to Infuse setting, and call the D
ELIVERY
T
IME
screen.
3.
Use the
number
buttons to type in the Delivery Time setting. Maximum and minimum times are calculated
for the syringe you are using. When ready, press the
Enter
button to call the T
IME
B
ETWEEN
S
TARTS
screen.
4.
Use the
number
buttons to set the Time Between Starts value. Time must be greater than delivery time and
less than or equal to 24 hours. When ready, press the
Enter
button to accept the setting go to the B
EGIN
I
NFUSION
screen.
5.
All settings are made, and the pump pauses at the B
EGIN
I
NFUSION
screen. Confirm all settings are correct.
You may now accept the settings and start the infusion, or use the four menu buttons to modify the settings.
6.
Prior to beginning the infusion, press <BOLUS> to initiate the priming sequence (page 21).
7.
After confirming all settings are correct, press Start to begin intermittent volume / time infusion.