Page 65
External DC Power Option
The infusion pump can also be run by an optional external DC power supply.
The external DC power supply must supply Input 6 – 14 VDC, 7 Watts.
This optional DC power supply is connected through a terminal on the rear of the pump.
The connection can be used for either an external battery or other low voltage DC input.
)
Use only
Medex
approved cabling and power supplies. Refer to Service Manual and pump labeling for this
information
.
Warning!
External Power
must be
IEC 601-1
Certified
Any external power source connected to the DC jack must be IEC 601-1 certified for
medical equipment Type CF and meet Safety Class II requirements. Connecting any
external power source to the
Medfusion 3500
pumps creates a medical system.
Therefore, the user is responsible for compliance with IEC 601-1 standards. Refer all
questions to the Medex Technical Service Department.