M
EDEX
Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 31
Recall Last Settings
The
Medfusion 3500
will
remember the settings used to run the last infusion indefinitely (or at least 10 years). This
means you can reuse the same delivery settings to repeat an identical dosage of medication or fluids to a patient.
1.
You may reuse the last settings by selecting R
ECALL
L
AST
S
ETTINGS
from the S
ELECT THE
M
ODE
screen. (If
you don’t see it, press the
More
button and search for it.)
When selected, you must then press the
Enter
button to confirm data for each screen that applies for the last
infusion. This includes all settings and options. The actual screens that appear depend on how your pump is
configured and which features are enabled or disabled.
2.
After reviewing these previously used settings, the pump pauses.
3.
If you wish to clear the TVD and PVD or PDD, press the <CLEAR Totals> button
4.
Prior to beginning the infusion, press <BOLUS> to initiate the priming sequence (page 21).
5.
Be sure all pump settings are correct before starting infusion.
TVD and PVD /
PDD Retention
The pump also retains TVD and PVD or PDD from the last infusion. If you do not want
to use these values, you must manually clear them before beginning a new infusion.