M
EDEX
Medfusion 3500
Medex Operation Manual
Artwork GD000608 Revision 5
Page 6
Understanding Front Panel Indicators
Below is a table of function indicators and their effects while using the pump.
Indicator
Pump Off
Pump Paused
Pump Delivering
−
AC Line Indicator
ON when AC powered.
ON when AC powered.
ON when AC powered.
–
Battery
Indicator
ON when charging.
ON when charging.
Flashing when battery in use.
ON when charging.
Flashing when battery in use.
–
Lock
Indicator
OFF
OFF
ON when keypad is locked.
−
Infusing Indicator
OFF
OFF
Flashing from right to left.
What Indicators Mean
Below is a table of front panel indicators and what each of them means:
Indicator
What it means
−
Alarm Indicators
The A
LARM
indicators (yellow or red) are on whenever the pump is in an alarm condition.
The specific details of each alarm are covered in the Alarm section later in this manual.
-
AC Line Indicator
The AC
L
INE
indicator (green) is ON whenever the pump is connected to “mains” line
power. It is OFF when the pump is not connected to an active AC line.
– Battery Indicator
The B
ATTERY
indicator (green) blinks ON&OFF whenever the pump is operating on
internal battery power, and remains ON when battery is charging.
– Lock Indicator
The L
OCK
indicator (
±
) tells you the pump has been locked into its current operational
mode.
While this indicator is lighted, the keypad is locked and no changes can be made to
settings. Attempting to stop or change an infusion while locked will result in an alarm and
an advisory message.
−
Infusing Indicator
The Infusing indicators are 3 green lights, which illuminate in sequence right to left when
the pump is infusing.
During intermittent delivery mode, a single Infusing indicator lights during the time
between infusions.