VR2250
Page 88 of 131
10. RECORDINGS
10.1. START A RECORDING
The recording function is available for
ANALYSER
and
AUX
rotary switch position.
As you can read in the paragraphs 7.2.1 and 8.2, a recording can be started manually or
automatically. Therefore, after setting all the parameters
and leaving the Menu
, the
instrument will start to record:
MANUALLY: the recording will start when the Instruments time reaches the “00”
seconds value after pressing
START/STOP
.
AUTOMATICALLY:
If the operator has pressed START/STOP
the instrument will
remain in stand-by until the date and time previously set, then the recording will start.
While
if the operator doesn’t press START/STOP the recording will never start.
CAUTION
For recordings
we recommend to use the external power supply
adapter
(optional code MAR####) even the instrument allows the operator
to perform a recording using internal batteries.
If you press Start a recording without the external power supply adapter (optional code
MAR####) the instrument will display a warning message “
No ext supply
”. Press
START
key again to run the recording or press
ESC
to quit.
If during a recording the external power supply adapter (optional code MAR####) is de-
energised, the instrument will continue the recording using the internal battery power until
the batteries are exhausted (the data stored until the definitive turning off won’t get lost).
For this we recommend you
ALWAYS insert a new set of batteries before a long
recording
.
The instrument uses sophisticated algorithms to prolong the battery life. Particularly:
The instrument switches OFF the backlight automatically after 5 seconds.
If the Battery level is too low the Backlight function will be disabled.
If the instrument is just displaying in real time (and the external power supply is not
connected), after about 5 minutes from the last key pressure or switch rotation the
instrument turns off automatically ("AUTOPOWER OFF" function).
If the instrument is recording or is measuring energy (and the external power supply is
not connected), after about 5 minutes from the last key pressure or switch rotation the
instrument starts a special procedure to save the batteries ("ECONOMY MODE"): the
instrument keeps recording but the display is turned off.
Before starting a recording the operator should first evaluate the state of the equipment,
decide what to record and set the instrument accordingly.