6.2.3. Results of "TMR" mode
At the end of the test if
the Insulation
resistance is
lower
than R
MAX
(see
Table2)
and the instrument
generated the
Nominal test Voltage
,
the instrument emits a
double sound signal
indicating the positive
outcome of the test
and displays one
screen similar to the
screen alongside.
M
Ω
05.06.01
Insulation Resistance
1.07
M
Ω
514V 15s
TMR 500V TIME:60s
Voltage during the Test
Duration of the Test
Test mode
Setting Time
FUNC
VNOM
↑
↓
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).
If the Insulation
resistance is
higher
than R
MAX
(see
Table2
),
the instrument emits a
double sound signal
at
the end of the test
indicating the positive
outcome of the test
and displays one screen
similar to the screen
alongside.
M
Ω
05.06.01
Maximum resistance value
which can be measured
(999
Ω
is displayed if a
rated voltage of 500V was
selected. See Table2).
> 999
M
Ω
523V 15s
TMR 500V TIME:60s
The symbol
">"
means that
the resistance value is
higher than R
MAX
.
Test
duration
FUNC
VNOM
↑
↓
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).