VR2250
Page 34 of 131
This screen is displayed
when the phase
conductor has been
reverseded with the
neutral one.
The instrument does not
perform the test.
Exchange the red cable
with the black one.
Repeat the test
RCD 05.06.01
- - -
ms
FRQ=50.0Hz Ut= ---V
VP-N=231V VP-PE= 0V
CHANGE P-N
x1 30mA
⎯
50V
Phase and Neutral
Conductors are reversed.
FUNC IdN RCD UL
This screen is displayed
when the phase
conductor has been
exchanged with the
Protection Conductors.
The instrument does not
perform the test.
Reverse the phase to
earth connection in the
plug or exchange the
red cable with the green
one.
RCD 05.06.01
- - -
ms
FRQ=50.0Hz Ut= ---V
VP-N= 2V VP-PE= 230V
CHANGE P-PE
x1 30mA
⎯
50V
Phase and Protection
Conductors are reversed.
FUNC IdN RCD UL
This screen is displayed
when in a 230V Phase
to Phase System the
black conductor was
reversed with respect to
the green one.
The instrument does not
perform the test.
Reverse the black and
green conductors.
RCD 05.06.01
- - -
ms
FRQ=50.0Hz Ut= ---V
VP-N=130V VP-PE= 227V
CHANGE N-PE
x1 30mA
⎯
50V
Neutral and Protection
Conductors are reversed.
FUNC IdN RCD UL