1133.12B1/B3/F1/F3
GA113312EN02
39
4.
Positioning a patient weighing up to 135 kg (300 lb)
4.1 Approved accessories
Injury hazard!
The accessories which are mounted may not exceed the maximum length as specified in the sketch.
For further accessories, please refer to the accessories list in chapter VIII.
4.2 Operating table not immobilized (UNLOCK)
4.2.1 General
Hazard of operating table tipping over!
If the patient is not positioned correctly, then the
operating table may tip over when extending the
operating table castors (UNLOCK).
The patient must be positioned according to Fig.
V-1 and the operating must be set to a height
less than 940 mm before extending the operat-
ing table castors (UNLOCK). The position below
940 mm is attained when the warning sign is not
seen.
NO
YES
Fig. V-8: Marking for moving patients weighing up to
135 kg
V. Patient positioning