1133.12B1/B3/F1/F3
GA113312EN02
47
VII. Technical specifications
1.
Operating table adjustments
Length without accessories:
866 mm
(34.1 in.)
Length incl. accessories
1130.67xx / 1131.31xx / 1133.53xx:
2090 mm
(82.3 in.)
Width:
530 mm
(20.9 in.)
Width across side rails:
max. 580 mm
(22.8 in.)
Height (without pad) 1133.02.12XX
594 - 1056 mm
(23.4 - 41.6 in.)
Weight:
approx. 312 kg
(688 lb.)
Trendelenburg:
30°
Reverse Trendelenburg:
30°
Lateral tilt left/right:
20°
Back plate up:
80°
Back plate down:
40°
Upper back plate up:
90°
(manual adjustment)
Upper back plate down:
45°
(manual adjustment)
Leg plate up:
10°
Leg plate down:
90°
Longitudinal shift:
230 mm
Flex position:
40°
Reverse Trendelenburg
20°
Back plate down
20°
Beach chair:
Trendelenburg
25°
Back plate up
80°
Leg plate down
70°
Max. patient weight:
450 kg
(1000 lb.)
2.
Electrical specifications
Nominal voltage:
AC 100/110-115/127
200/220/230-240V
Frequency:
50/60 Hz
Power consumption:
400 VA max.
Class:II.
Applied part:
Type B
The enclosure leakage current meets the requirements of the patient leakage current for CF conditions
according to IEC 60601-1, IP X4* IPS ** SELV*** DC 24V
Operating mode:
int. 10 min ON / 20 min OFF
For patient weights from 225 kg to 450 kg: 2 min ON / 120 min OFF
Explosion protection
AP for operation on battery power
When connected to the mains, the operating table is not
approved for use in areas subject to explosion hazard
(AP-M)!
Rechargeable batteries:
Type A512C / 15SR
2 Batteriens 12 V / 15 Ah
* IPX 4 =
protection against ingress of liquids
**
IPS = internal power source
***
SELV =
safety extra low voltage
VII. Technical specifications