z)
0-position
Resets the lateral tilt, Trendelenburg/reverse
Trendelenburg, back plate and leg plate ad-
justments to the horizontal position. If the
function button z is pressed, the operating
table top will first reset the adjustments in the
following order: lateral tilt, Trendelenburg/re-
verse Trendelenburg, back plate. An acoustic
signal will indicate that these adjustments have
been reset. Release the function button z and
press once again to reset the leg plates to
the horizontal position; if the leg plates have
been adjusted individually, they will be posi-
tioned in parallel to one another.
Notes:
Manual adjustment functions such as those for the
upper back plate, head plate and leg plate abduc-
tion will not be affected by the IR remote control
(and, as a consequence, neither by the automatic
0-position function).
The transmission power for the IR remote control
ist sufficient to operate the draped operating table
even from a greater distance. If the desired adjust-
ment is not executed by the operating table, even
though a function button is depressed completely,
then you should re-aim the hand control or move to
another position in the OR. In such cases the re-
ceiver is probably being blocked by a person’s body.
2.2 Override system = emergency operation
2.2.1 Override control panel
In case of a malfunction or if the hand control is defec-
tive, you can control the operating table functions via
the control panel (13) on the upper column casing (emer-
gency operation).
Note:
Always press the ON button together with the de-
sired function button.
Note:
If the operating table batteries are discharged, the
override system will only work when the operating
table is connected to the mains.
Note:
When using the override control panel, the collision
protection feature is not activated. Avoid extreme
adjustments - the combination of maximum Tren-
delenburg/reverse Trendelenburg position and maxi-
mum lateral tilt.
13
III. General description
Fig.III-4:
Override control panel