1133.12B1/B3/F1/F3
GA113312EN02
16
14
14
III. General description
2.5 Control via foot switch 1009.81DX
(optional)
Connect the foot switch to one of the two sockets (14)
at the column head.
Three functions each can be activated using the follow-
ing foot switch versions:
1009.81D0
Foot switch with the functions:
height, Trendelenburg/rev. Trendelen-
burg, back plate
1009.81D1
Foot switch with the functions:
height, Trendelenburg/rev. Trendelen-
burg, lateral tilt
1009.81D2
Foot switch with the functions:
height, Trendelenburg/rev. Trendelen-
burg, leg plates
1009.81D3
Foot switch with the functions:
height, Trendelenburg/rev. Trendelen-
burg, longitudinal shift
Note:
Observe the operating instructions for the foot
switch.
2.6 „Neurolock“ function
If requested, the MAQUET service personnel can set
the „Neurolock“ function at the operating table.
This function avoids adjustments of the table position
due to unintentional actuation of the buttons on the hand
control.
a+b) The „Neurolock“ function can be unlocked by
pressing simultaneously the buttons „table top
down“ (a) and „table top up“ (b) on the control
module. The operating table can now be adjusted.
If no further buttons on the control module are
actuated for approx. 8 seconds, then the
„Neurolock“ function will be activated again. Ad-
justing the operating table again is only possible
after actuating simultaneously the „table top
down“ (a) and „table top up“ (b) function buttons.
Note:
The „table top down“ and „table top up“ functions
cannot be actuated simultaneously using the foot
switch 1009.81X0.
Use an IR hand control or a corded hand control to
unlock the „Neurolock“ function.
Fig.III-12: Foot switch
Fig.III-13: „Neurolock“ function