Part No. 20717 Rev-01
2014-07
vi
Use Outside of Product Labeling
• The INOmax DS
IR
must only be used in accordance with the indications, usage,
contraindications, warnings and precautions described in the INOMAX (nitric
oxide) drug package inserts and labeling. Refer to this material prior to use.
• The manufacturer does not recommend that the INOmax DS
IR
be utilized with
helium/oxygen mixtures in any situation. The INOmax DS
IR
is intended to deliver
INOMAX therapy gas only in conjunction with the delivery of air and oxygen. The
use of helium/oxygen mixtures will lead to over-delivery of INOMAX which may
lead to interruption of therapy.
• The use of devices which radiate high-intensity electrical fields may affect the
operation of the INOmax DS
IR
. Constant surveillance of all monitoring and life
support equipment is mandatory whenever interfering devices are in operation on
or near a patient.
• Do not connect items which are not specified as part of the system.
• The approved patient population for the INOmax DS
IR
, as specified in the drug
labeling for INOMAX (nitric oxide) for inhalation, is limited to neonates. The
INOmax DS
IR
is not intended to be used in other patient populations.
Transport
• If the INOmax DS
IR
or INOblender is to be used in a transport vehicle, they should
be affixed to the transport mounting post (part number 10009), which is part of
the transport mounting bracket assembly (part number 50041).
• The transport mounting post and/or the transport mounting bracket assembly
should be secured to the transport isolette/transport gurney in a manner which
will secure the INOmax DS
IR
/INOblender.
Troubleshooting or Calibrating
• If an alarm occurs, safeguard the patient first before troubleshooting or repair
procedures.
• Use caution when troubleshooting the INOmax DS
IR
delivery system while
in use for a patient. When possible replace the unit in question and perform
troubleshooting procedure once the unit is removed from the patient.
• If changing an NO sensor while delivering NO to a patient, install the NO sensor
only when the NO high range calibration screen is displayed otherwise there is a
risk that the system will shut down.
• INOMAX can be administered during the sensor calibration process. However,
inspired gases are not monitored and gas monitoring alarms are disabled.
• Loss of communication between the INOmax DS
IR
and the INOMAX cylinder for
more than one hour will result in interruption of INOMAX delivery.
WARNING:
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...