Part No. 20717 Rev-01
2014-07
5-1
5/ Alarms
General alarm information
A listing of alarm messages is provided at the end of
this section.
All alarms have audible tones and visual messages.
In the event of a total power failure or a main
alarm speaker failure, a secondary audible alarm
circuit activates, providing a continuous buzzing
tone that can not be silenced (see page 5-6,
"Continuous Audible Tone").
Note: Status information will not be displayed
during alarm conditions. Once the alarm
clears the status information will be displayed.
High and low-priority alarms
The INOmax DS
IR
has both high and low-alarm priorities.
High-priority alarms are accompanied with a red
flashing Alarm Silence button.
Low-priority alarm conditions will display a continuous
yellow Alarm Silence button.
High and low-priority alarm messages are displayed
in fields one and two (see Figure 5-1) with the most
recent message shown in field one.
Field 2 is used for status information such as “Running
on Battery” and “Patient Info. Incomplete”.
Frequency Description
Comment
High
Priority
400 Hz
10-pulse group
Repeats after 10 sec.
if not silenced.
Low
Priority
400 Hz
1 pulse
Repeats after 40 sec.
if not silenced.
The following table provides the audible alarm tone
information for high and low-priority alarms.
Figure 5-2 Alarm Active Screen
1. Alarm Silence Button
2. Alarm Help Button
3. Violated Monitored Value Limit
2
1
3
Alarm silencing
Pushing the Alarm Silence button
will silence high-priority alarms for
120 seconds (time will count down
to zero). When a new alarm condition occurs, the
audible alarm becomes active again.
A low-priority alarm event is permanently silenced
when the Alarm Silence button is pressed. When a
new low-alarm condition occurs, the audible alarm
becomes active again.
Alarm messages remain displayed during the
alarm silence period as long as the alarm condition
is active.
Field 1
Field 2
To activate on-screen alarm help, press
the Alarm Help Button next to the Alarm
Silence button.
Caution:
• Any alarm setpoint adjustments made will not be maintained when system power is cycled.
• Default values will be used following a complete power loss (no AC main power and depleted
battery).
Figure 5-1 Text Message Area Showing Fields
one and two.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...