Part No. 20717 Rev-01
2014-07
5-10
High Priority Alarms
Alarm
Possible Cause
Recommended Action
8. Delivery Failure
a. Over-delivery of INOMAX.
- (CALCULATED dose is
>200% of set dose) AND
(CALCULATED dose is
> set dose + 10 PPM) for
12 consecutive seconds,
or
- CALCULATED dose >
100 for 12 seconds
INOMAX delivery has been interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
b. The internal NO flow sensor
temperature or measured
flow rate is out of range.
Other internal error
detected (a temporary
internal communication
error that has been
detected).
If Delivery Failure Alarm Occurs:
1. Use one of the back-up INOMAX delivery
options
2. Turn INOmax DS
IR
to standby, then restart.
3. Once device is ready to restart therapy:
4. Turn integrated back up off (if used) and then
set the INOMAX dose.
5. Confirm delivery and check alarms.
Contact Technical Support.
9. Delivery Stopped
a. MONITORED NO > 100
ppm for at least 12 seconds
(once MONITORED NO
drops below 100 ppm for 12
seconds, alarm will resolve
and INOMAX delivery will
restart).
INOMAX delivery has been interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
b. Drug past expiry date
or drug concentration
mismatch.
Replace INOMAX cylinder.
c. INOmeter may have failed.
Replace INOMAX cylinder.
d. INOMAX cylinder valve is
closed.
Open INOMAX cylinder valve.
e. INOMAX cylinder is not
detected.
Replace INOMAX cylinder.
Contact Technical Support.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...