Part No. 20717 Rev-01
2014-07
4-8
INOblender Test Using the INOmax DS
IR
to Analyze Output
WARNING:
• The purge procedure must be followed to help ensure NO
2
is purged from the
system before the manual resuscitator bag is connected to the patient.
• The manual bag should be squeezed repeatedly during use to avoid NO
2
building
up in the bag.
• If the bag is not squeezed repeatedly while delivering INOMAX, the bag should
be removed from the patient and the bag purge procedure performed before
continuing.
•
The INOblender should be upright when setting the oxygen flowrate for accurate
setting.
• Do not use pneumatically powered nebulizers with the INOblender. This will result
in significant over delivery of INOMAX in excess of 80 parts per million (ppm).
- The INOblender outlet pressure has been validated for use up to 400 millibar
(5.8 psig) pressure. The amount of back-pressure generated by pneumatic
nebulizers is significantly greater (20-30 psig) and will result in over delivery
of INOMAX in excess of 80 ppm. The user adjusted dose setting on the
INOblender will not correlate with, or have an effect on the actual delivered
dose.
- In addition, the INOblender flowmeter is not back-pressure compensated
and will display a lower flow rate than actual when pressure is applied to the
outlet.
Caution:
•
When not in use, the oxygen flowmeter and the INOMAX cylinder valve should be turned off.
• When the INOblender is used with an oxygen/air blender:
- The specification for INOMAX delivery when using the INOblender with 100% oxygen is
+/- 20% of setting or two ppm (whichever is greater). The use of 100% oxygen at 50 psig
is the labeled specification for the INOblender.
- A user may determine that some clinical conditions may necessitate the use of an
oxygen/air blender with the INOblender to achieve FiO
2
levels less than 100%.
- Using oxygen/air mixtures (21% to 95% v/v) will reduce the delivered NO concentration
by up to 10% of setting or 1 ppm (whichever is greater) compared to using 100% oxygen
alone, resulting in a cumulative error up to +/- 30% of setting or three ppm (whichever is
greater).
•
Refer to the manufacturer's procedures when using the manual resuscitator bag.
•
When finished, turn the NO cylinder valve OFF and allow the oxygen to flow until the NO
pressure gauge reads zero, then turn the oxygen flow OFF.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...