Part No. 20717 Rev-01
2014-07
9-7
1. Verify that the O
2
flowmeter is set to 10 L/min.
2. Set the INOMAX dose to 40 ppm, allow
values to stabilize.
3. Compare the INOmax DS
IR
monitor values to
the values in the table.
4. Turn INOMAX dose to zero.
Set INOMAX Dose
40 ppm
Acceptable NO
Value
35-45 ppm
Acceptable NO
2
Value
< 1.5 ppm
FiO
2
95% ± 3 %
Performance Test
Note: Allow 2-3 minutes for monitored values
to stabilize. If a monitored value is
outside the range indicated, do a high
range calibration for that sensor.
1. Verify that the O
2
flowmeter is set to 10 L/min.
2. Turn the integrated backup INOMAX delivery to
ON (250 mL/min.).
Verify "Backup ON" alarm occurs.
3. Allow monitored values to stabilize (may take
up to 3 minutes) and make sure the NO and
NO
2
readings are within the following ranges:
NO = 14-26 ppm
NO
2
≤ 1.0 ppm
4. Turn the integrated backup INOMAX delivery
to OFF.
Integrated Pneumatic Backup INOMAX Delivery Test
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...