Part No. 20717 Rev-01
2014-07
iv
High Frequency Oscillatory and Jet Ventilator Circuits
• Some high frequency ventilator circuits require a one-way valve to prevent high
NO delivery.
• Place the Bunnell Life Pulse in Standby prior to suctioning the patient to avoid
NO delivery transiently exceeding the set dose by up to 30 ppm. Press ENTER to
reestablish ventilation as soon as the catheter is removed from the airway. This
will limit the extent of over delivery above the NO set dose.
• Do not use dose settings above 40 ppm with the Acutronics Fabian HFO
ventilator. Bidirectional flow through the Injector Module may cause over-delivery
which can lead to measured NO values greater than 100 ppm
Maintenance
• For continued protection against hazard, replace the fuses only with the correct
fuse type and rating.
• Handle and dispose of sensors according to facility biohazard policies. Do not
incinerate.
• If the Injector Module has been used in the wet/humidified part of the breathing
circuit, it should be sterilized between each patient use.
• INOmax DS
IR
should only be connected to RS-232 ports that have:
- 4 kV input to output isolation
- 4 kV input to mains isolation, and
- an internal “reference voltage” “U” (as defined in section 20.3 of IEC60601-
1 ed. 2) of less than or equal to 50 VDC or 50 VRMS and dielectric isolation
certified in accordance with IEC 60601-1. Interface cabling must not go outside of
the room (e.g., into walls where potential isolation issues could exist). Adherence
to the above provides compliance to clause 20.3 “Value of test Voltage” in edition
2 and clause(s) 8.5.4 “Working Voltage” and Clause 8.8.3 “Dielectric Strength" in
edition 3.
• RS232 cable must be shielded. The RS232 cable shield shall have a minimum
of 90% coverage. The shield shall only be connected at one end of the cable to
minimize noise induced by ground currents.
Manually Bagging a Patient with an Injector Module
• The hyperinflation bag will, under some conditions, contain NO
2
in excess of
one ppm. Use of large tidal volume breaths may expose the patients to the NO
2
present in the bag for part of the breath. In general, if the inspiratory flow rate
induced by the manual ventilation does not exceed the fresh gas flow rate,
the patient should not be exposed to the concentrations of NO
2
present in the
hyperinflation bag.
• Adult and infant hyperinflation bags generate more NO
2
when used at lower
minute ventilation. If use of the bag is interrupted (for example to adjust the
tracheal tube), before resuming ventilation of the patient, the user should squeeze
the bag several times to empty residual gas from the bag.
• Because of the potential for inhalation of excessive concentrations of NO
2
, and
the difficulty in monitoring the peak inhaled NO
2
concentrations, ventilation with
a hyperinflation bag or self-inflating bag is intended only for short term use.
• The monitoring system within the INOmax DS
IR
will not detect generation of NO
2
within the hyperinflation bag or self-inflating bag devices and the alarms for
excessive NO
2
cannot warn of NO
2
produced within the manual bag system.
WARNING:
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...