Part No. 20717 Rev-01
2014-07
1-15
The INOmax DS
IR
has an interface using infrared (IR) technology which allows the INOmax DS
IR
to
communicate with the INOmeter
(which is mounted to each INOMAX cylinder). The INOmax DS
IR
checks the
INOMAX cylinder for the correct expiration date and cylinder concentration. The INOmax DS
IR
also transmits
a confirmed patient identifier to the
INOmeter on any open INOMAX cylinder.
The INOmax DS
IR
cart (PN 10018) has a cover (see Figure 1-7, ) with an infrared transceiver mounted
directly above each INOMAX cylinder. When INOMAX cylinders are loaded, communication will take
place between the INOmax DS
IR
and the INOmeter (see Figure 1-7, ) after the boot up phase of the
INOmax DS
IR
is complete. A cylinder icon will be displayed on the main screen when an INOMAX cylinder is
recognized by the INOmax DS
IR
(see “Loading INOMAX Cylinders onto the INOmax DS
IR
Cart”, page 1-17).
1
2
Infrared Communication between the INOMAX Cylinders and
the INOmax DS
IR
IR Communication Interference
The INOmax DS
IR
transceiver is located under the cart cover and should be protected from outside IR
sources. The INOmax DS
IR
cart was designed to protect the INOmeter from external light/IR energy sources.
The INOmax DS
IR
transceiver transmits via a 30 degree transmission cone projecting towards the floor (see
dotted lines in Figure 1-7
). The specifications of the IR beam call for it to have a range of 20 cm (7.9 in). Based
on these specifications it sould not affect other devices in the vicinity of the INOmax DS
IR
.
The INOmeter uses a lower energy source which results in a lower IR beam range than the INOmax DS
IR
cart.
The INOmeter does not transmit IR signals unless it is mounted on the INOmax DS
IR
cart.
If there is interference with the INOmax DS
IR
/INOmeter communication, the cylinder icon on the user screen
will not be displayed and a “Cylinder Not Detected” alarm will activate if there is a set INOMAX dose.
If IR communication interference occurs, we recommend taking the following actions:
• Move the external IR source
• Move the INOmax DS
IR
cart to reduce the external IR source in the area of the INOmeter
• Shield the INOmeter from the suspect IR source
If the actions listed above do not remedy this issue, the transport regulator/cap assembly (PN 10022) may
be utilized.
WARNING:
Loss of communication between the INOmax DS
IR
and the INOmeter for more
than one hour will result in interruption of INOMAX delivery.
Caution:
Nothing should be placed between the INOmeter and the cart to which it is attached.
Caution:
A strong magnetic field could affect the ability of the
INOmeter to detect if the cylinder valve is
opened or closed. This may affect the ability of the INOmax DS
IR
to detect the position (open
or closed) of the cylinder valve.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...