Part No. 20717 Rev-01
2014-07
5-7
High Priority Alarms
Alarm
Possible Cause
Recommended Action
2. Low NO alarm
All of these actions can be performed while
delivering INOMAX to the patient:
a. The Low NO alarm setting
may be inappropriately set.
Confirm low NO alarm limit is set appropriately.
b. Circuit setup incorrect.
1. Check circuit for correct setup.
2. Confirm water trap bottle, water separator
cartridge, NO/N
2
injector tube and patient gas
sample line are in place.
c. Loss of NO delivery.
If loss of NO delivery is suspected, manually
ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least 5 L/min ventilator gas flow in
patient circuit).
d. The Patient Gas Sample
line may be disconnected.
1. Check circuit for correct setup.
2. Confirm water trap bottle, water separator
cartridge and patient gas sample line are in
place.
e. The NO calibration may
have drifted.
1. Perform low calibration
2. Perform high calibration.
f. The NO sensor may not be
properly seated.
Confirm the O-rings on the sensor and sensor
cover are correctly seated and the sensor cover
is fully closed.
g. Injector module may not be
functioning properly.
INOMAX delivery will be interrupted.
Manually ventilate patient with the INOblender
or
turn integrated pneumatic backup delivery ON
(verify at least five L/min ventilator gas flow in
patient circuit).
1. Change injector module.
2. Change injector module cable.
“Injector Module Failure” alarm will occur and
NO delivery will be interrupted. The injector
module failure alarm will resolve when
injector module cable is reconnected.
Contact Technical Support.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...