Part No. 20717 Rev-01
2014-07
2-2
1. Confirm the water trap bottle and water
separator cartridge are in place .
Connect the patient gas sample line with
Nafion to the sample line inlet port on the
front of the INOmax DS
IR
.
2. Connect one end of the injector module
electrical cable to the injector module and
the other end into the front of the
INOmax DS
IR
.
• Line up the red dot on both the connector
and the injector module before inserting
the connector (see inset detail).
Connect one end of the INOMAX injector
tube to the injector module and the other end
into the front panel of the INOmax DS
IR
.
Check cables and hoses for signs of wear and
damage.
WARNING:
Be certain all cables and
hoses are positioned to
help prevent damaging or
occluding them.
Note:
• It is recommended to disinfect or
sterilize the injector module prior to
initial setup.
• To remove this type of connector, the
knurled sleeve on the connector
must be pulled outward before
removing the connector from the
injector module or the front panel.
Initial connections
1b
1b
1a
1a
2a
2b
2b
2c
2a
2c
2a
2b
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...