Part No. 20717 Rev-01
2014-07
iii
Blue arrow denotes required user action.
WARNING:
Integrated Pneumatic Backup
• The integrated pneumatic backup is intended for short term use when the electronic
delivery system fails until a replacement NO delivery device can be brought to the
bedside.
• The integrated pneumatic backup delivers a variable concentration of NO to the
patient depending on the ventilator flow being used.
• When using the integrated pneumatic backup with breathing circuit gas flows of
5 L/min, the delivered NO dose will be approximately 40 ppm. Breathing circuit gas
flows less than 5 L/min will deliver an NO dose greater than 40 ppm.
• The integrated pneumatic backup (250 mL/min.) should not be used with the
Bunnell Life Pulse as ventilator flow rates are normally below the recommended
ventilator flows.
Changing Cylinders
• Always secure a cylinder when not using it.
• Never lift a cylinder by its valve or valve protection cap or by using a chain, sling or
magnet.
• Never drop a cylinder.
• Never use a hammer, pry or wedge to loosen a valve or protection cap. The valve
and protection cap should operate by hand.
• Never let oil, grease or other combustible come in contact with a cylinder or valve.
• Never remove or deface cylinder labeling or markings.
• Never modify equipment without first contacting the manufacturer.
• Never use an adaptor to connect a cylinder to the system.
• Never use equipment not designed to use INOMAX mixtures.
• Never attempt to repair a leak on a cylinder valve or its safety relief device.
• Never operate equipment that is leaking.
• Never ship a leaking cylinder.
• Never store cylinders;
- Where damage can result from the elements, such as standing water or
temperatures over 125 degrees F.
- Where they can experience extreme low temperatures.
- Where they can contact corrosive substances.
- Where they can be cut or abraded by an object.
- Next to a walkway, elevator or platform edge.
- Unless they are properly secured.
WARNING:
Warnings tell the user about dangerous conditions that can cause injury to
the operator or the patient if you do not obey all of the instructions in this
manual.
Caution:
Cautions tell the user about how to properly use the equipment and conditions
that could cause damage to the equipment.
Read and obey all warnings and cautions.
Note:
Notes provide clarification or supplemental information.
Содержание INOMAX DSir Plus
Страница 10: ...Part No 20717 Rev 01 2014 07 viii Intentionally left blank...
Страница 11: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 12: ...Part No 20717 Rev 01 2014 07 1 General Information General Information...
Страница 17: ...Part No 20717 Rev 01 2014 07 1 5 Intentionally left blank...
Страница 39: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 40: ...Part No 20717 Rev 01 2014 07 2 Automated Pre Use Checkout Automated Pre Use Checkout...
Страница 51: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 52: ...Part No 20717 Rev 01 2014 07 3 Patient Application Patient Application...
Страница 83: ...Part No 20717 Rev 01 2014 07 3 31 Intentionally left blank...
Страница 103: ...Part No 20717 Rev 01 2014 07 3 51 Intentionally left blank...
Страница 106: ...Part No 20717 Rev 01 2014 07 3 54 Intentionally left blank...
Страница 107: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 108: ...Part No 20717 Rev 01 2014 07 4 Transport Transport...
Страница 128: ...Part No 20717 Rev 01 2014 07 4 20 Intentionally left blank...
Страница 129: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 130: ...Part No 20717 Rev 01 2014 07 5 Alarms Alarms...
Страница 148: ...Part No 20717 Rev 01 2014 07 5 18 Intentionally left blank...
Страница 149: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 150: ...Part No 20717 Rev 01 2014 07 6 Calibration Calibration...
Страница 165: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 166: ...Part No 20717 Rev 01 2014 07 7 Maintenance Maintenance...
Страница 180: ...Part No 20717 Rev 01 2014 07 7 14 Intentionally left blank...
Страница 181: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 182: ...Part No 20717 Rev 01 2014 07 8 Product Specifications Product Specifications...
Страница 195: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 196: ...Part No 20717 Rev 01 2014 07 Appendix 9 Appendix...
Страница 205: ...Part No 20003 Rev 01 Part No 20717 Rev 01 2014 07 Intentionally left blank...
Страница 206: ...Part No 20717 Rev 01 2014 07 Mallinckrodt Manufacturing LLC 6603 Femrite Drive Madison WI 53718 6801 USA 877 566 9466...