40
41-621.1 HART
®
to Modbus
®
Adaptor Operating Manual
12.4.12
Create a Transmit Macro that will send U03OL?$0D and click on the Mx button to the
right of the macro definition. This command requests the value for Level Offset in the
attached device. Note that the 03 in the macro represents the Poll Address of the Model
706 device.
12.4.13
Verify that the Terminal application receives a response from the HMA each time the Mx
button is clicked, and that there are no communication errors being reported. The
Receive buffer section should have the same number of responses as the Transmit buffer
section.
12.4.14
Verify that the value after the OL in the response matches the Level Offset value shown
on the local display of the Model 706 device. The value displayed is shown as an integer
number to conform to the command requirements, but actually represents the Level
Offset multiplied by 10. For example, a Level Offset of 1.5 inches will be displayed in the
command response as 0015. The value returned by the command may vary from the
value shown on the local display by a value of 1 due to rounding. The value returned will
be in terms of Level Units.