Originally
Adopted:
07/13/2006
Revision #:
1.1
Revision Effective Date: 09/08/2006
Subject Sample Correlation Failure
This exception occurs when the subject did not provide two similar valid breath samples. This
could be a result of improper breath sample delivery or residual alcohol in the subject's upper
respiratory tract. The instrument will display
“Sample Corr. Fail”
and print
“*Invalid Test –
Subject Sample Correlation Failure”
. The test is not complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should consider
checking the subject’s mouth. 2) Conduct another pre-test wait/observation period of at least 15
minutes. 3) Consider re-instructing the subject on how to provide a proper breath sample. 4) Use a
new mouthpiece and restart the test. 5) Continue the test until the subject properly submits to or
refuses the test. 6) Any observations made by the Operator during the testing sequence may be
recorded in the comments section.
Sequence Aborted by Operator
This exception occurs when the operator aborts the testing sequence. The instrument will display
“Sequence Aborted”
and print
“*Invalid Test – Sequence Aborted by Operator”
. The test is
not complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) If the test was terminated in
error, the test may be restarted by the operator. 2) Any observations made by the Operator during
the testing sequence may be recorded in the comments section.
Range Exceeded
This exception occurs when the test result was higher than 0.60 %BAC. This could be due to a
high concentration of residual alcohol in the subject’s mouth or upper respiratory tract from recent
consumption of alcohol or having vomited or regurgitated liquid from the stomach into the mouth.
The instrument will display
“Range Exceeded”
and print
“*Invalid Sample – Range Exceeded”
.
The test is not complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should consider
checking the subject’s mouth. 2) Conduct another pre-test wait/observation period of at least 15
minutes. 3) Use a new mouthpiece and restart the test. 4) Continue the test until the subject
properly submits to or refuses the test. 5) Any observations made by the Operator during the
testing sequence may be recorded in the comments section.
Control Out of Tolerance
This exception occurs when the result obtained by the instrument during the testing of the external
ethanol breath standard is out of tolerance. The instrument will display
“Control Outside
Tolerance”
and print
“*Invalid Test – Control Out of Tolerance”
. The test is not complete and
can be restarted.
ALL PRINTED COPIES ARE UNCONTROLLED AND MAY NOT BE COMPLETE