O
REGON
S
TATE
P
OLICE
I
MPLIED
C
ONSENT
P
ROGRAM
I
NTOXILYZER
®
8000
O
PERATOR
’
S
G
UIDE
4.0
S
TEP
T
HREE
-
R
EVIEWING THE
T
EST
R
EPORT
Originally
Adopted:
07/13/2006
Revision #:
1.1
Revision Effective Date: 09/08/2006
1. If the subject submitted to the test, the instrument will print the numerical result obtained from
the analysis in the Test Result location. The result will be printed as a two-digit number
followed by the measurement units (e.g.,
“0.14 %BAC”
).
2. If the subject refused the test and the operator pushes the “Start Test” button or the “R” key on
the instrument keyboard, the instrument will print
“Refused”
in the Test Result location.
3. If the instrument detected an interfering substance in the person’s breath sample, the
instrument will print
“Interferent Detected”
in the Test Result location.
Complete Test Reports should be reviewed by the operator. This way, the operator is certain the
testing sequence was successful.
The Test Result value reported by the instrument on the Completed
Test Report will be lower value of the two subject samples provided truncated to two-digits
(e.g.,
0.089 %BAC will be reported as 0.08 %BAC).
Incomplete Test Reports
Incomplete Test Reports can be differentiated from Complete Test Reports in several different ways:
1. The Test Result will not contain any numerical information, a refusal, or an interfering
substance detected statement.
2. The instrument will print a statement clarifying the message contained in the Test Result
location immediately underneath the Test Result location in the section titled “Additional
Comments”. This clarification message will begin with the statement “*Invalid Test” or “*Invalid
Sample” followed by additional descriptive information.
3. If appropriate, the instrument will also print a “Suggested Corrective Action” statement to be
read and acted on by the operator.
Incomplete Test Reports should be reviewed by the operator. The Suggested Corrective Action
statement on the Test Report will often contain the necessary information for the operator to re-initiate
the testing sequence and obtain a Complete Test Report.
Common Exception Messages:
The Intoxilyzer 8000 breath alcohol analyzer system is a micro-processor controlled breath alcohol
testing device that includes the capability of monitoring its internal systems during a test sequence. As
a result, the instrument is able to alert the operator whenever a condition that might affect the result of
a breath test occurs. The Intoxilyzer 8000 instrument is designed to only allow a test result to be
reported when no exceptions have occurred during the entire testing sequence. Otherwise, the
instrument will alert the operator that an exception has occurred by displaying and printing one or
more of the following messages: