The Continued Observation Period will be considered incomplete if:
1. The operator has reason to believe that the subject has taken something by mouth (drinking,
smoking, eating, taking medication, etc.), vomited, or regurgitated liquid from the stomach into
mouth; or
2. The subject, through some willful act, refuses the breath test; or
3. The operator (by pressing the “Start Test” button or pressing the “A” key on the instrument
keyboard) or the instrument aborts the testing sequence.
In the event of a discontinued or incomplete Observation Period, the operator should abort the testing
sequence, document the reason for halting the testing sequence, and restart the Pre-Test
Observation Period. The operator can then retest the subject.
Air Blank
After waiting for approximately two-minutes, the instrument will perform another
“Air Blank”
. The
instrument will display the
“Air Blank”
message and turn on the instrument’s internal purge pump for
approximately 20 seconds.
Air Blank
Air Blank
0.000
This will purge the instrument’s sample chamber and once again check the operational environment
for possible contaminants. If all the parameters are met, the instrument will proceed to the next step.
If there is a problem, the instrument will abort the test and display a message of
“Ambient Fail”
on
the display and print
“*Invalid Test – Check Ambient Conditions”
on the Test Report.