Residual Alcohol Present
This exception occurs when residual alcohol was detected in the subject's mouth or upper
respiratory tract. This could be the result of recent consumption of alcohol or having vomited or
regurgitated liquid from the stomach into the mouth. The instrument will display
“Residual
Alcohol Present”
and print
“*Invalid Sample – Residual Alcohol Present”
. The test is not
complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should consider
checking the subject’s mouth. 2) Conduct another pre-test wait/observation period of at least 15
minutes. 3) Use a new mouthpiece and restart the test. 4) Continue the test until the subject
properly submits to or refuses the test. 5) Any observations made by the Operator during the
testing sequence may be recorded in the comments section.
RFI Detected
This exception occurs when a radio or cellular phone transmission is detected near the instrument.
The instrument will display
“RFI Detected”
and print
“*Invalid Test – RFI Detected”
. The test is
not complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should consider
removing the radio frequency transmission source from the vicinity of the instrument.
2) Use a new mouthpiece and restart the test. 3) If the instrument gives an additional exception
message, consider placing the instrument out of service by placing an “Intoxilyzer Out Of Service”
form on the instrument and reporting the problem to the OSP Regional Dispatch Center at 1-800-
292-6517. 4) Leave the instrument power on and consider locating an alternative testing location.
5) Any observations made by the Operator during the testing sequence may be recorded in the
comments section.
Unstable Signal
This exception occurs when the instrument could not obtain a stable reference during the testing
sequence. The instrument will display
“Unstable Signal”
and print
“*Invalid Test – Unstable
Signal”
. The test is not complete and can be restarted.
In addition, the following suggested corrective action is provided: 1) The operator should check for
any obvious reason(s) for the condition (e.g., did the subject blow at the wrong time?) 2) Use a
new mouthpiece and restart the test. 3) If the instrument gives an additional exception message,
consider placing the instrument out of service by placing an “Intoxilyzer Out Of Service” form on
the instrument and reporting the problem to the OSP Regional Dispatch Center at 1-800-292-
6517. 4) Leave the instrument power on and consider locating an alternative testing location. 6)
Any observations made by the Operator during the testing sequence may be recorded in the
comments section.