Air Blank 

The instrument will initiate the testing sequence by performing an air blank.  The instrument display 
will display the 

“Air Blank”

 message and turn on the instrument’s internal purge pump for 

approximately 20 seconds.   
 
 
 

Air Blank 
0.000

Air Blank 

Air Blank 
0.000

 
 
 
 
The instrument will draw in room (“ambient”) air to purge the sample chamber and breath path of any 
possible alcohol or other contaminants. If all the parameters are met, the instrument will proceed to 
the next step.  If there is a problem, the instrument will abort the test and display a message of 

“Ambient Fail”

 on the display and print 

“*Invalid Test – Check Ambient Conditions”

 on the Test 

Report (refer to Section 4.0 – Reviewing the Test Report). 
 

Diagnostic Routine 

Following the successful completion of the 

“Air Blank”

, the instrument will perform a complete 

diagnostic check on its components and operational standards. 
 
 
 

Diagnostics 

Diagnostics 

ƒƒƒƒƒƒƒƒƒƒƒƒƒƒƒƒ

 

 
 
 
 
If no operational malfunctions are found, the instrument will display a message of 

“Diagnostics 

Pass”

 and then proceed to the next step.  

 
 
 

Diagnostics 
Pass

 
 
 
 
If any malfunctions in the operation of the instrument are detected, the instrument will abort the test 
displaying a message on the display. 

 

ALL PRINTED COPIES ARE UNCONTROLLED AND MAY NOT BE COMPLETE

 

Page 16 of 54

 

 

Содержание INTOXILYZER 8000

Страница 1: ...Oregon State Police Forensic Services Division OPERATION GUIDE INTOXILYZER 8000 OPERATOR S GUIDE...

Страница 2: ......

Страница 3: ...Press the Start Test button 9 Selecting a Test Method 9 Entering Permit Data Manually 10 Entering an Operator PIN 11 Scanning a Driver s License 11 Entering Driver s License Data Manually 12 Reviewing...

Страница 4: ...eports 41 On site Reprinting of Test Reports 41 Entering Permit Data Manually 41 Entering an Operator PIN 42 Selection of a Test Report for Reprinting 43 6 0 CONDUCTING A JAIL RELEASE TEST 44 Conducti...

Страница 5: ...onducted in Oregon by promoting and maintaining consistency and quality among Intoxilyzer 8000 operators This Guide will serve as a good reference for the operation of the Intoxilyzer 8000 instrument...

Страница 6: ...ecognizes efforts of the following individuals and groups for their contributions to this manual Department of Public Safety Standards and Training Oregon State Police Drug Recognition Program Implied...

Страница 7: ...urposes of this guide the System consists of two 2 primary components 1 the Intoxilyzer 8000 instrument and 2 the external printer Because the instrument is micro processor controlled certain steps sh...

Страница 8: ...rying handle 8 Flexible heated breath tube 9 External standard connection 10 Mouthpiece storage area Figure 1c Intoxilyzer 8000 Parts Controls and Indicators back 11 Main AC power switch rocker switch...

Страница 9: ...that the battery power switch be placed in the OFF DOWN position first followed by the main AC power switch To operate the instrument using the on board back up battery system place the main AC power...

Страница 10: ...tes prior to submitting the breath sample This is to ensure there is no possibility of residual alcohol or mouth contamination causing a result that does not reflect the subject s true blood alcohol c...

Страница 11: ...ining the person s eligibility for release from a jail detox or other detention facility The instrument will acknowledge the operator s selection to conduct a jail release test by updating the instrum...

Страница 12: ...operator should type in his or her operator permit number DPSST number As the operator enters information on the keyboard the information will be displayed on the second line of the instrument s displ...

Страница 13: ...cond line of the instrument s display The instrument will validate the information provided by the operator and acknowledge that a valid PIN has been entered by proceeding to the Scan Subject License...

Страница 14: ...line of the instrument s display Generally name information is entered in a uniform format similar to that used by the LEDS Law Enforcement Data System reporting system last name first name middle na...

Страница 15: ...f subject identification number information by proceeding to the Enter State of Issue message Enter State of Issue The operator should type the two 2 letter designation for the jurisdiction responsibl...

Страница 16: ...After displaying each entry the operator must accept the information by pressing the Enter key on the instrument s keyboard Aborting a Test Sequence During the testing sequence an operator may choose...

Страница 17: ...nding on the amount of time necessary for the operator to complete the data entry section of preparing to conduct a breath test the instrument may display the Ready in 00 XX message with a timer that...

Страница 18: ...e instrument will proceed to the next step If there is a problem the instrument will abort the test and display a message of Ambient Fail on the display and print Invalid Test Check Ambient Conditions...

Страница 19: ...rogress bar as a series of rectangular characters i e and then begin scrolling the message Please blow into mouthpiece to activate tone across the top line of the instrument s display At this time the...

Страница 20: ...t the Y key on the instrument keyboard to document a refused test or select the N key on the instrument keyboard to document an Invalid Test Deficient Sample on the Test Report After approximately one...

Страница 21: ...ith sufficient force to activate the audible tone during the three minute request period The instrument will display the No Sample Given message accompanied by a low high alternating tone sounding int...

Страница 22: ...ination that the test subject does not have the physical capacity to complete a breath test and is not a suitable candidate to continue with the breath testing process If however the operator has dete...

Страница 23: ...ol or mouth contamination causing a result that does not reflect the subject s true blood alcohol concentration The Intoxilyzer does monitor the breath sample throughout the entire delivery and will a...

Страница 24: ...Observation Period the operator should abort the testing sequence document the reason for halting the testing sequence and restart the Pre Test Observation Period The operator can then retest the subj...

Страница 25: ...Please Blow message As the subject continues to blow several indicator marks will appear across the display When a period is displayed following the Please blow message making the message read Please...

Страница 26: ...ubject refuses the test and the operator pushes the Start Test button or the R key on the instrument keyboard to indicate to the instrument the operator s request that the instrument document the refu...

Страница 27: ...e If the operator has properly instructed the test subject on how to provide the breath sample and the person has provided a deliberate attempt to meet the requirements of the test the operator may ch...

Страница 28: ...the instrument will proceed to the next step If there is a problem the instrument will abort the test and display a message of Purge Fail on the display and print Invalid Test Purge Fail on the Test R...

Страница 29: ...Test Report Comment Lines Immediately prior to printing the Test Report the operator will be given the opportunity to enter comments into the Comments section of the report This information will then...

Страница 30: ...nd space for remarks or handwritten information deemed pertinent by the operator A Completed Test Report is one which indicates a breath test result a refusal or the presence of an interfering substan...

Страница 31: ...g the Start Test button during the breath sampling period i e the Please Blow phase The instrument will display the Subject Test Refused message NOTE It is suggested that operators use the instrument...

Страница 32: ...roblem the instrument will abort the test and display a message of Sample Corr Fail on the display and print Invalid Test Subject Sample Correlation Failure on the Test Report Operators can refer to O...

Страница 33: ...rth license number and state of issue 5 The test method steps the intermediary results of each step the status of each step and the time each step was initiated 6 The final result of the test 7 Any ad...

Страница 34: ...fusal or an interfering substance detected statement 2 The instrument will print a statement clarifying the message contained in the Test Result location immediately underneath the Test Result locatio...

Страница 35: ...The instrument will display Improper Sample and print Invalid Test Improper Sample The test is not complete and can be restarted In addition the following suggested corrective action is provided 1 Th...

Страница 36: ...Refused This exception occurs when the operator terminates the testing sequence because the subject refused to continue The instrument will display Subject Test Refused and print Subject Test Refused...

Страница 37: ...er removing the radio frequency transmission source from the vicinity of the instrument 2 Use a new mouthpiece and restart the test 3 If the instrument gives an additional exception message consider p...

Страница 38: ...n be restarted In addition the following suggested corrective action is provided 1 If the test was terminated in error the test may be restarted by the operator 2 Any observations made by the Operator...

Страница 39: ...uld check for any obvious reason for the condition e g did the subject blow at the wrong time 2 Use a new mouthpiece and restart the test 3 If the instrument gives an additional exception message cons...

Страница 40: ...to perform all steps identified in a Suggested Corrective Action statement is misplaced Other Messages Displayed by the Instrument Tank Pressure Low Change Standard Soon If the instrument detects tha...

Страница 41: ...re the instrument aborts the testing sequence Incorrect PIN Permit Expired If the operator s permit has expired the instrument will display the Permit Expired message The operator should contact the O...

Страница 42: ...Intoxilyzer 8000 so that the instrument cannot be used by operators When the instrument is disabled the instrument will display Disabled Mode and the date and time Disabled Mode 07 12 2006 14 00 PDT O...

Страница 43: ...isplay the following message Scan Operator Permit or Press Enter When the instrument displays the Scan Operator Permit or Press Enter message the operator may use the instrument s internal bar code sc...

Страница 44: ...alid permit number and PIN combination is entered Entering an Operator PIN Upon successfully scanning an operator permit or the manual entry of a valid permit number and operator name the instrument w...

Страница 45: ...e reprint process It should be noted that only Test Reports generated by the operator will be available to the operator for subsequent reprinting Test Reports generated by other operators are not acce...

Страница 46: ...XX message From the Ready in 00 XX display the instrument will display a timer that will count down the period of time needed for the instrument to be ready to conduct a breath test This period of ti...

Страница 47: ...acceptable breath sample is the instructions that are given to the test subject The following instructions are recommended I WANT YOU TO TAKE A DEEP BREATH AND BLOW INTO THE INSTRUMENT WITH ONE LONG C...

Страница 48: ...he instrument displaying a period at the end of the Please Blow message The instrument will terminate the breath sampling phase before the end of the three minute period if any of the following condit...

Страница 49: ...e Deficient Sample message By selecting a Deficient Sample the operator has made the determination that the test subject does not have the physical capacity to complete a breath test and is not a suit...

Страница 50: ...e the instrument will record a Refused test sequence rather than a Sequence Aborted test Breath Test Refusal During a testing sequence an operator may decide that a subject has through some willful ac...

Страница 51: ...be sure to perform the suggested correction action that is provided with the exception message The dispatcher will inquire as to whether the operator has performed this step prior to placing the inst...

Страница 52: ...OXILYZER 8000 OPERATOR S GUIDE APPENDIX A OPERATOR S CHECKLIST Originally Adopted 07 13 2006 Revision 1 1 Revision Effective Date 09 08 2006 APPENDIX A OPERATOR S CHECKLIST ALL PRINTED COPIES ARE UNCO...

Страница 53: ...Language v6 compatible printer may be used Parts Controls and Indicators Figure 1 below explains the parts controls and indicators of the external printer Brother HL 5250DN If a different model of pri...

Страница 54: ...ncels the print operation in progress 6 GO BUTTON Wake up Error recovery Form feed Printer LED Indications The following table explains the various LED indications provided by the Brother HL 5250DN pr...

Страница 55: ...ear Them If a paper jam occurs the printer will flash the yellow Paper LED WARNING After you have just used the printer some internal parts of the printer will be extremely hot When you open the front...

Страница 56: ...he printer 13 Push down the blue lock lever and remove the toner cartridge from the drum unit Clear the jammed paper if present NOTE Handle toner cartridge carefully If toner contacts hands or clothes...

Страница 57: ......

Отзывы: