StoreX
Series
STX_V078.doc
74
CMa,
23.11.2009
Stop Shaker
RS
sp
1913
cr
OK
cr lf
Read Shovel Plate Sensor
RD
sp
1812
cr
x
cr lf
Read Transfer Station Plate Sensor
RD
sp
1813
cr
x
cr lf
Read 2
nd
Transfer Station Sensor
RD
sp
1807
cr
x
cr lf
Read Swap Station position
RD
sp
1912
cr
x
cr lf
Read Actual Temperature (1/10º Celsius)
RD DM982 cr ttttt
1
cr lf
Read Set Temperature value (1/10º C; e.g. 370
= 37.0º Celsius)
RD DM890
cr
sssss
2
cr lf
Set operation temperature (1/10º C)
WR DM890 sp ttttt
cr
OK cr lf
Read Actual Humidity (1/10% RH)
RD DM983 cr ttttt
cr lf
Read Set Temperature value (1/10% RH; e.g.
900 = 90.0% RH)
RD DM893
cr
sssss
cr lf
Set operation humidity (1/10% RH)
WR DM893 sp ttttt
cr
OK cr lf
Read Actual CO2 level (1/100% vol.)
RD DM984 cr ttttt
cr lf
Read Set CO2 level (1/100% vol. 500 = 5.0%
CO2)
RD DM894
cr
sssss
cr lf
Set operation CO2 level (1/100% vol.)
WR DM894 sp ttttt
cr
OK cr lf
Read Actual N2 level (1/100% vol.)
RD DM985 cr ttttt
cr lf
Read Set N2 level (1/100% vol. 9000 = 90.0%
N2)
RD DM895
cr
sssss
cr lf
Set operation N2 level (1/100% vol.)
WR DM895 sp ttttt
cr
OK cr lf
The stacker slot is selected by sending m=1..2. The handler is positioned towards
the specified slot. Note that stacker levels are numbered starting at the bottom
level upwards. Bottom level is 1.
The Activate Handler-Command initializes the handling. Use the Activate Handler
Command after cold-start or after sending a Reset Command. The Activate
Handler Command should not be used as an ‘on-off’ function.
Functions marked with *
)
can only be used with selected stacker slots. The value
of the number of stackers is stored in DM29. The value of the number of levels is
stored in DM25. When using the Position z-Lift for BCR reading command the ST
1910 command has to be sent only once. Once set, the handler is positioned by
simply writing the carrousel and level position into DM0 and DM5.
The Activate Plate Sensor command is used in HT-units only. In HT units the
Shovel Plate Sensor is deactivated by default. The Activate Plate Sensor
command has to be sent prior to reading the sensor status (ST 1911). This
command will turn the sensor on for ~2 seconds. The sensor can be read during
this period of time. After turning the sensor on wait approximately 0.1 of a second
before reading the sensor with the RD 1812 command.
The Error-Flag is set when the handling cannot finish a movement in time. Check
the Error-Flag by sending “RD 1814” when the Ready-Bit does not become ‘1’.
The Plate Ready flag (1815) is set when the system has cleared the plate from
the Transfer Station on executing an Import command or when a plate is placed
on the Transfer Station during an Export command. This Plate Ready flag allows
increased access speeds. The read Plate Ready (RD 1815) returns '1' until the
read Ready flag (RD 1915) is set to '1'.
1
t = actual value 5 characters (Word=16Bit)
2
s = set value 5 characters (Word=16Bit)