StoreX
Series
STX_V078.doc
18
CMa,
23.11.2009
3.5.3
N2 Nitrogen Option
As StoreX incubators have to cope with frequent accesses they use a high
quality N2 controller. The nitrogen measurement is based on an oxygen
measurement. The N2 option therefore assumes that oxygen is displaced by
nitrogen.
The N2 Option of StoreX incubators consists of
•
Gassing conditioner
•
Gassing Controller
•
Measurement Cell
Principle of Operation
A zirconia electrolyte disc with electrodes on both sides has the ability to pass
oxygen ions at high temperature.
By attaching a cap with an aperture on one side (cathode) of the disc, the gas
diffusion is limited and the saturated current is observed. It is termed "limiting
current". The limiting current is proportional to the ambient oxygen concentration.
Specifications
Measuring Range
75..99.9 % Vol.
Input Pressure
2
8
bar typ.
bar max.
Gas Flow
~10 l/min max.
Gas Consumption @ 95% N2, 20°C
<__
~__
~__
l/h standby
l/h 2-min.-access cycles
l/h 30sec.-access cycles
Accuracy ±0.5
%
FS
Stability
±0.5 % FS over 12 month
Repeatability
±1 % at full scale