In order to adjust the Transfer Station in the Y-direction (perpendicular to the
back-plane) use the screws (3) and (4) and slide the Transfer Station into
position then tighten screws again. The Height of the Transfer Plate is adjusted
by opening screw (5) and lifting or lowering the Transfer Plate. Again, tighten
screw (5) after finishing the height adjustment.
WARNING !
The Gate will move when running the STXFER utility program
Do not put hands into the Gate opening.
There is a PC-run utility program to ease the adjustment of the Transfer Station.
Start the program „STXFER.EXE“ on the STX Utility Diskette. After Startup the
program will ask you the enter the comport to the StoreX.
If you do not know the port number to the StoreX we recommend you run this
program on a separate PC or laptop.
A warning will appear. Read the warning carefully. If you understood the warning
and understand this set-up procedure type ‘Y’ to continue. If you have questions
contact Liconic or your integrator.
Transferplate
Vertical Adjustment
Screw (5)
Y-Adjustment
Screw (3), (4)
X-Adjustment
Screw (1), (2)