StoreX
Series
STX_V078.doc
31
CMa,
23.11.2009
Instruments with cooling options should not be operated immediately after
transportation. Leave instrument turned off for at least 12 hours.
The picture below shows another possibility for arranging StoreX units. This
example depicts the combination of two StandAlone units and one BenchTop
unit. This arrangement allows maximum usage of valuable laboratory space. For
easy integration and short access paths the Gate of BenchTop units is located
near the bottom of the unit while the Gate of Stand Alone units is located near the
top of the instrument.
Front Panel
Allow min.
30mm clearance
above ground
Front Door
Ventilation Slits
Allow ~ 10mm
clearance between
units