Two-stage pressure regulators are available from gas suppliers or laboratory
accessories suppliers. The picture above shows an example of an installation.
There is a wide variety of different models available. Your equipment may look
different.
The gas inlet of the StoreX is located on the back of the instrument. Attach the
CO2 low pressure hose securely to the fitting. Always use the security nut. Never
operate the unit without the security nut being tightly fit.
Use 4x6mm PUR, PVC, Neoprene or Nylon tube for the low pressure connection.
Do not use Silicon or natural rubber tubing. Make the low pressure tube long
enough to allow a minimum of movement of the StoreX for cleaning and
maintenance work. For longer length of tubing, copper or other metal tubing
should be considered. Foresee flexible tubing next to the instrument.
Be prepared for the following gas consumption values. These values may vary
significantly over time since they depend on the instruments use.
6.4.5
Specific Advice for CO2 Carbon Dioxide
Principals of Operation
CO2 is mainly used in cell-based assays. CO2 prevents changes in the PH-value
of the probes. Although CO2 is not toxic several precautions must be taken in
order to ensure safe operation.
WARNING !
Improper use of CO2 or improper installation of the CO2 operated equipment
may cause death
Read CO2 safety information and installation instructions (See CO2
Installation) carefully. If you are not familiar with the handling of CO2 and / or
safety precautions contact trained personnel.
Security nut
Low pressure pipe
StoreX gas inlet