7.8 Power-On
System
The Instruments power is activated by the incubator’s main switch. The
instrument is turned on by pressing the power switch located at the front of the
instrument. After power-on the climate portion of the instrument is activated
automatically. An alarm will sound. If the green light of the Power Switch does not
turn on, check the Main Switch at the back of the instrument.
The alarm indicates that the handling is not activated. In order to activate the
handling, the red Start Switch has to be pressed. The Start Switch will cause the
handling to initialize. Take note whether the Gate - if opened- is closed after
pressing the Start Switch. Pressing the Start Switch is equal to sending an
external Activate Command through the Communication Port.
It is normal for some condensation to occur during warm-up. This condensation
will disappear once the internal climate is stabilized. The settling time may take
several hours.
The temperature inside the Climate Chamber is displayed on the Temperature
Display. The Temperature Display may not be used to set Temperature. The
Operating Temperature can only be set through the Communication Port.
Power Switch
Display Temperature
Indicator Light
Display Humidity
Display Option 1
(e.g.CO2)
Display Option 2
Multi-Function Switch