7.2 Front
Door
The Front Door opens on the left side of the instrument and should be kept
closed at all times. A Magnetic Seal keeps the Front Door closed. When access
is necessary it should be for as short as possible periods.
In order to avoid condensation and to allow an even temperature distribution
inside the instrument, the Front Door is heated. Note: Leaving the Front Door
open for extended periods will influence the instruments temperature.
The Front Door is opened by pulling on the left edge and may be opened to an
angle of almost 180 degrees when standing alone. When placed next to another
StoreX unit the Front Door may be opened by approximately 100 degrees.
Regularly inspect sealing performance of the Front Door. The Magnetic Seal
must keep the Front Door firmly closed over its whole surface. When the Front
Door is out of alignment with the Front Shield it must be re-aligned by an
authorized person. Note that gaps between the Magnetic Seal and Front Shield
will cause undesired air exchange and therefore condensation and will degrade
the temperature distribution.
CAUTION !
Use left edge of Front Door to open Front Door
Keep Front Door closed when it is not being accessed
Keep Front Door Access to a minimum
Magnetic Seal must firmly close Front Door
Keep Magnetic Seal clean
WARNING !
It is dangerous to leave the Front Door open over long periods of time.
Avoid excessive condensation on the glass door
Front Door
Pull left edge of Front Door
Magnetic Sealing
Front Shield