Rotation / Translation
These fields display the calculated rotation (in degrees) and translation (in millimeters).
Statistics
The statistics provide information on the best fit. The first line shows the maximum 3D
deviation and the ID for which the deviation was determined. The second line indicates the
minimum 3D deviation and the corresponding ID, followed by the mean 3D deviation and the
3D standard deviation.
By pressing the
Back
button, you can access the previous pages of the wizard and make
modifications. For example, you can exclude the ID that shows the maximum 3D deviation
from the calculation by returning to the page
Select the pass points
and deselecting that
specific point. Alternatively, you can also measure the point again.
Press
Finish
. A new group named “Setup1” is created in the data explorer. All existing data
sets are moved to this group. Another group named “Setup2” is also created and activated.
Note
Do not move data sets in or out of the “Setup” groups. Data sets in one of these groups were
measured in a specific tracker position and should remain in that group. However, you can still
create subgroups and delete or move data sets within the “Setup” groups.
15.9.6 Invoking Move Station again
If the wizard is started anew, the page
Select the pass points
(and not
Measure the pass points
of the initial position) will open after pressing the button
Next
because the reference points for
the current tracker position have already been acquired during the last execution of the
wizard. This also applies if you restart the T-Scan Collect application and load a file that
involved working with the wizard.