11. REMOTE CONTROL (LV 5838 ONLY)
F•1
●
Configuring the LV 5838
To set the remote control connector, use the ETHER&REMOTE tab in the system settings.
[Reference] ETHER&REMOTE tab → See section 5.2, “Configuring the Ethernet and Remote Settings (LV
5838
only).”
→
SYSTEM SETUP
→
NEXT TAB
→
Figure 11-2 ETHER&REMOTE tab
●
Control
The input connectors respond to active-low signals. Do not apply negative voltages or
voltages that 5 V. The active-low signal must be stable for at least 350 ms. After that,
wait at least 1 second before applying the next signal.
Figure 11-3 Remote control timing 1
After a setting is made, it may take about 3 seconds for the operation to finish. If you
configure subsequent settings before the initial operation finishes, only the last setting will
take effect. All settings in between will be discarded. (In the following example, control 2 will
be discarded.)
Figure 11-4 Remote control timing 2