5. SYSTEM SETTINGS
36
●
Trigger: OFF / Remote / Timecode / CM
Set the automatic loudness measurement.
OFF:
Automatic measurement is disabled. Use the START/PAUSE key to start
and stop a measurement. Use the CLEAR key to clear a measurement.
Remote:
Measurement start, stop, and clear are executed through the remote
control connector. On the ETHER&REMOTE tab in the system settings,
you must set Select to Recall and 5838_02 or 02. You cannot
select this on the LV 5837.
Timicode: Measurement start and stop are executed on the basis of the time codes
embedded in the input signals. Enter values for StartTime and EndTime.
You must set TIMECODE SETTING on the GENERAL tab in the system
settings to some setting other than OFF.
CM:
Measurement start, stop, and clear and storage to USB memory are
executed at the following times. Enter a value for StopTime.
If you select CM, CM appears in the upper right of the screen. Press START/PAUSE,
and then apply the material to measure. At this point, the indication in the upper right of
the screen changes to
CM
.
Figure 5-9 CM explanation
a) Loudness measurement is cleared 1 second after the detection of audio (1 kHz).
b) Loudness measurement starts 0.1 seconds after the detection of no audio.
MEAS
appears in the upper right of the screen.
c) Loudness measurement stops when the time specified by StopTime elapses after the
detection of no audio. If a USB memory device is connected, the loudness data is
automatically saved to the device.
CM
appears in the upper right of the screen.
You can measure multiple materials in succession by applying subsequent materials.
No audio
Audio (1 kHz)
No audio
Audio (body)
No audio
Audio (1 kHz)
No audio
a
b
c
a
b
USO
RESTRITO
Содержание LV 5837
Страница 1: ...LV 5838 LV 5837 AUDIO MONITOR INSTRUCTION MANUAL USO RESTRITO...
Страница 98: ...12 ETHERNET CONTROL LV 5838 ONLY 90 26 AUD_VALIDITY_ERR AUDIO Validity error detection USO RESTRITO...
Страница 101: ...12 ETHERNET CONTROL LV 5838 ONLY 92 Country or region Time Zone Kiribati Line Islands Tokelau 14 00 USO RESTRITO...