Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 14-4
Menu selection
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chap
ter 14
14.3 Auto on off menu item 3
Maintenance interval:
Time interval for the reminder to carry out
system maintenance
6, 12, 24 months
12 months
Change pre-filter:
Time interval for the reminder to change the pre-filter
0…8 weeks
8 weeks
MB sampling
Time interval for the reminder to carry out microbiological
sampling (germ sample)
1…12 months
1 month
Timer reset
After maintenance work, a filter change or a germ sample,
the date for the next action to be carried out is updated
•
Maintenance interval
•
Change pre-filter
•
MB sampling
Cell/cond. raw water
0,05…0,50 1/cm
adjustment
Parameter
Function
Range
Default setting
NOTICE
With hard water alarm, the discharge is automatically coupled to the degree of hardness.
Hardness degree °dH
ppm CaCO3
WCF
0–5
0–89
60 %
1 °dH (German hardness degree)
= 1.78 °f (French hardness degree)
= 1.253 °e (English hardness degree)
= 1.042 gpg CaCO
3
= 17.8 ppm CaCO
3
6–10
90–178
50 %
11–15
179–267
40 %
16–20
268–356
30 %
21–25
357–440
25 %
>25
>440
20 %
When the main switch is set to 1, dialysis mode is carried out conti-
nuously until the operator manually switches the system to off (switch
position 0) or to standy mode (switch position 2).
Two dialysis operating modes per day can be programmed in this
program menu, these are then initiated automatically.
By selecting the weekdays, the programmed times can be viewed or
changed (depending on the pre-selection >Display< or >Entry<).
The pre-set program times can be reset by selecting >Clear all values<.
1 or 2 dialysis modes can be entered for each day which determine the
start and end of dialysis mode through real-time entries (hh:mm).
Auto on/off
Monday (1)
Tuesday (1)
Wednesday (1)
Thursday (1)
Friday (1)
Saturday (1)
Sunday (1)
Clear all values?
<<
OK
Weekday (1)
Start: xx h xx min
Stop:
xx h xx min
<<
OK
Weekday (2)
Start: xx h xx min
Stop:
xx h xx min
<<
OK