Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 8-2
Starting / Shutting down
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chapter 8
8.4
Putting the machine out of operation
If the system is to be shut down for a longer period, it is recommended that the system be treated with
preservative.
After a longer period of standstill, the system is put into operation again as described under
chapter 3 “Work prior to commissioning”.
8.5
Return and disposal
Figure 8-1:
Battery removal
In accordance with legislative regulations, Lauer Membran Wassertechnik GmbH offers to take back
systems it supplies and dispose of these as stipulated by legislation.
WARNING
Please note
part 1, chapter 8.6 “Technical memo for PRESERVATION with sodium
metabisulphite” on preserving the system.
Electrical medical devices are subject to the Guideline for used electric and electronic appli-
ances 2002/96 dated January 27, 2003, and have to be disposed of in an environmentally
friendly manner.
The medical device contains a battery. Used batteries should be recycled. The battery is
located on the CPU (PCB with display) in the switch cabinet door and can be lifted off the
PCB using a screwdriver.
Carry out the following steps:
1. Switch the main switch to “0”
2. Screw off the EMC cover
3. Replace the battery
Battery (1 pc. / item no. 40410)
CPU
Screwdriver