Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 15-6
Faults / Causes / Correction
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chap
ter 15
When Y9 is triggered, the signal must come from
FISCAL 3
DI: for error:
DI 19
Delay:
9 sec
Self acknowledging:
yes
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
System continues to run, discharge according to dialysis mode parameters, menu 2.17
Emergency mode RO I:
System continues to run, discharge according to dialysis mode parameters, menu 2.17
Emergency mode RO II:
System continues to run
Standby mode:
System continues to run
Disinfection:
System continues to run
Cleaning:
System continues to run
HotRO I/ HotRO I+II
System continues to run
Pressure switch PSAH 1 has been triggered
DI: for error:
DI 14 = 0
Delay:
No
Self acknowledging:
yes
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
System OFF
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
System OFF
Level has dropped below minimum pressure
at pressure switch PSAL2.;
only active if M3 on
DI: for error:
DI 15 = 1
Delay:
15 sec
Self acknowledging:
yes
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
RO continues to run with M1 + M2, pump M3 off, after self-acknowledging M3 on again
Emergency mode RO I:
No function
Emergency mode RO II:
See Alarm/Error 20
Standby mode:
RO continues to run with M1 + M2, pump M3 off, after self-acknowledging M3 on again
Disinfection:
RO continues to run with M1 + M2, pump M3 off, after self-acknowledging M3 on again
Cleaning:
RO continues to run with M1 + M2, pump M3 off, after self-acknowledging M3 on again
HotRO I/ HotRO I+II
No function
Level has dropped below minimum pressure
at pressure switch PSAL2 three times within
one minute.;
only active if M3 on
DI: for error:
DI 15 = 1
Delay:
5 sec
Self acknowledging:
--
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
No function
Emergency mode RO II:
See Alarm/Error 20
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
No function
wd. dd.mm.yy ss:mm
Error 15
PSAL2 prim.pres ROII
Menu Reset
wd. dd.mm.yy ss:mm
Alarm 15
PSAL2 prim.pres ROII
Menu Reset
wd. dd.mm.yy ss:mm
Alarm 14
Exc.press.ring PSAH1
Menu
wd. dd.mm.yy ss:mm
Alarm 13
FISCAL3 defective
Menu Reset