Rev. 4.00 December 2014
Software version 1.04
Part 1 • Page 15-10
Faults / Causes / Correction
Part 1 – Operating Instructions
EcoRO Dia I/II (HT)
Part 1 • Chap
ter 15
If Y9 is not triggered despite pulses from
FISCAL 3
DI: for error:
DI 19
Delay:
5 sec
Self acknowledging:
yes
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System continues to run
Emergency mode RO I:
System continues to run
Emergency mode RO II:
No function
Standby mode:
System continues to run
Disinfection:
System continues to run
Cleaning:
System continues to run
Hot RO I/ Hot RO I+II
System continues to run
The calculated amount of concentrate (WCF)
cannot be discharged through Y9/FISCAL3.
(no alarm 27 if concentrate > 2500 l/h)
DI: for error:
--
Delay:
5 min
Self acknowledging:
--
Reset+Rem. contr. ackn: yes
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
No function
Standby mode:
No function
Disinfection: No
function
Cleaning: No
function
HotRO I/ HotRO I+II
No function
No signal at analogue input CH 5 temperature
DI: for error:
--
Delay:
10 sec
Self acknowledging:
--
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System OFF
Emergency mode RO I:
System OFF
Emergency mode RO II:
No function
Standby mode:
System OFF
Disinfection: System
OFF
Cleaning: System
OFF
HotRO I/ HotRO I+II
System OFF
Level has dropped below limit value at pressure
switch PSAL4. Alarm will only be analysed when
pump M1 = on
DI: for error:
DI 20 = 1
Delay:
5 sec
Self acknowledging:
yes
Reset+Rem. contr. ackn: --
Operating phase
Initiating operating phase/action
Dialysis mode:
System continues to run
Emergency mode RO I:
System continues to run
Emergency mode RO II:
System continues to run
Standby mode:
1.
If via menu 4.13 PSAL 4 is active = 0 => no alarm message on the display
2.
If via menu 4.13 PSAL4 is active = 1 and ISS active ( if DI20 = 0 then Y30 close
3.
If menu 4.13 PSAL4 active = 1 and ISS operating time M4.11 expired ( Alarm message
on display and system continues to run
Disinfection:
System continues to run
Cleaning:
System continues to run
HotRO I/ HotRO I+II
System continues to run
wd. dd.mm.yy ss:mm
Alarm 26
SV Y9 leaks
Menu Reset
wd. dd.mm.yy ss:mm
Alarm 29
PSAL4 low press ring
Menu
wd. dd.mm.yy ss:mm
Error 28
Temp. meas. not OK
Menu
wd. dd.mm.yy ss:mm
Alarm 27
NV 4 conc. not OK
Menu Reset