Lake Shore Model 475 Gaussmeter User’s Manual
8-2
Service
8.3
FUSE DRAWER
The fuse drawer supplied with the Model 475 holds the instrument line fuses and line voltage selection module.
The drawer holds two 5
×
20 mm time delay fuses. It requires two good fuses of the same rating to operate safely.
Fuse ratings are different for different line voltages. Refer to Paragraph 8.5 for details.
EU_Fuse.eps
Figure 8-1. Fuse Drawer
8.4
LINE VOLTAGE SELECTION
Use the following procedure to change the instrument line voltage selector. Verify the fuse value whenever line voltage
is changed.
WARNING:
To avoid potentially lethal shocks, turn off controller and disconnect it from AC power before
performing this procedure.
1. Identify the line input assembly on the instrument rear panel. See Figure 8-2.
2. Turn the line power switch OFF (
O
).
3. Remove the instrument power cord.
4. With a small screwdriver, release the drawer holding the line voltage selector and fuse.
5. Slide out the removable plastic fuse holder from the drawer.
6. Rotate the fuse holder until the proper voltage indicator shows through the window.
7. Verify the proper fuse value.
8. Re-assemble the line input assembly in the reverse order.
9. Verify the voltage indicator in the window of the line input assembly.
10. Connect the instrument power cord.
11. Turn the line power switch On (
l
).
475_Power.bmp
Figure 8-2. Power Fuse Access
8.5
FUSE REPLACEMENT
Use the following procedure to remove and replace the line fuses.
WARNING:
To avoid potentially lethal shocks, turn off controller and disconnect it from AC power before
performing these procedures.
CAUTION:
For continued protection against fire hazard, replace only with the same fuse type and rating
specified for the line for the line voltage selected.
Содержание 475
Страница 16: ...Lake Shore Model 475 Gaussmeter User s Manual 1 6 Introduction 1 4 SAFETY SYMBOLS ...
Страница 25: ...Lake Shore Model 475 Gaussmeter User s Manual Background 2 9 Figure 2 8 Examples of PI Control ...
Страница 36: ...Lake Shore Model 475 Gaussmeter User s Manual 3 10 Installation This Page Intentionally Left Blank ...
Страница 68: ...Lake Shore Model 475 Gaussmeter User s Manual 4 32 Operation This Page Intentionally Left Blank ...
Страница 82: ...Lake Shore Model 475 Gaussmeter User s Manual 6 6 Remote Operation Figure 6 2 Model 475 Status System ...